Technical Specification

How-to Videos

Literature

Consumables

GastroCH4ECK® Gastrolyzer®

Direct sampling

The only breath hydrogen and methane device with direct breath testing allowing you to take a reading there and then, with instant results

Revolutionary Interface

Full-colour touchscreen with an easy-to-use interface and onscreen motivational exhalation guides

Breath bag sampling

Breath test samples can be captured with a breath bag and analysed at a later date or time.

Technical Specification

Technical spec icons for Nobreath website-01

Concentration range
CH4: 0-200ppm
 H2: 0-200ppm
O2: 14-23%

Technical Specifications icon

Accuracy
Resolution: 1ppm
Accuracy: ±10% of reading
Repeatability: <5% difference on consecutive readings

Technical spec icons for Nobreath website-06

Power
Mains power

Technical spec icons for Nobreath website-08

Operating temperature
15-35ᵒC

Technical spec icons for Nobreath website-11

Sensor operating life
CH4: 5 years
 H2: 2 years
O2: 2 years

Technical spec icons for Nobreath website-02

Display
Full-colour touchscreen

Technical Specifications icon

Accuracy
Resolution: 1ppm
Accuracy: ±10% of reading
Repeatability: <5% difference on consecutive readings

Technical spec icons for Nobreath website-12

Start-up time
≤2 minutes

Technical Specifications icon

Storage/transport temperature
0-40ᵒC

Technical Specifications icon

CO cross sensitivity
<4% (H2 only)

Technical spec icons for Nobreath website-03

Detection principle
Electrochemical sensor (O2 & H2)
Laser (CH4)

Technical Specifications icon

Accuracy
Resolution: 0.1%
Accuracy: ±10% of reading
Repeatability: <5% difference on consecutive readings

Technical spec icons for Nobreath website-07

Response time
≤45 seconds

Technical spec icons for Nobreath website-10

Operating humidity
912-1114mbar (Atmospheric ±10%)

Technical Specifications icon

Calibration frequency
Every 4 weeks

As easy as…

Direct breath test:

Breath bag test:

Literature

Gastro icons for website-02

Catalogue

Gastro icons for website-03

Datasheet

Gastro icons for website-01

Manual

Gastro icons for website-04

Infection Control
Maintenance Guidelines

Consumables

Gastrocheck mouthpiece

GastroCH4ECK® mouthpiece:

Specifically designed with the latest moisture removal and bacterial filtration to remove 99.9% of airborne bacteria from the patient’s breath1.

Facemask adaptor

The facemask adaptor is single-patient-use. A facemask is attached to one end and then the sampling system connects to the breath bags.

Moisture removal filter

For use with the breath bag sampling mode. These will last for approximately 150 samples before it needs to be changed.

GastroCH4ECK® breath bags:

Specifically designed to capture as much end-tidal air as possible, with a one-way mouthpiece to prevent loss of sample.

Facemask

Facemasks are single-patient use and
allow the patient to breathe normally
to produce a breath sample. There are
3 sizes: Large, Medium, and Small.

Calibration gas

The GastroCH4ECK® requires
calibration every 4 weeks using 100ppm H2,
100pm CH4, and 20.9% O2, provided as a kit or
replacement cylinder.

Breath bag
Face mask

References